ORIGINAL ARTICLE
Year : 2020  |  Volume : 43  |  Issue : 2  |  Page : 108-114

Calibration of MTS-N (LiF: Mg, Ti) chips using cesium-137 source at low doses for personnel dosimetry in diagnostic radiology


1 Department of Radiology, Medical Physics Unit, Federal Medical Centre Asaba, Asaba, Delta, Nigeria
2 Department of Radiation Biology, Radiotherapy and Radiodiagnosis, College of Medicine, University of Lagos, Lagos, Nigeria

Correspondence Address:
Mr. Akintayo Daniel Omojola
Department of Radiology, Medical Physics Unit, Federal Medical Centre Asaba, Asaba, Delta
Nigeria
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/rpe.RPE_6_20

Rights and Permissions

Thermoluminescent dosimeter (TLD) is still in use for many applications such as radiation protection, medical dosimetry, environmental research, and personnel dosimetry, with the overall aim of estimating radiation dose within a given medium or material. The aim of this study was to determine the coefficient of variation (CV) for thermoluminescent (TL) element within the same bar-coded slide and to establish calibration factors (CFs) at dose equivalent of 0.07 mm depth in tissue (Hp [0.07]) and dose equivalent of 10 mm depth in tissue (Hp [10]) for newly purchased TL elements alongside a new RadPro TLD manual reader and annealing oven. Annealed TL elements were taken to a Secondary Standard Dosimetry Laboratory (SSDL) for irradiation using a cesium-137 source at known doses (0.2–2 mGy). A RadPro Cube 400 manual TLD Reader was used to determine corresponding TL signal. The CV between two identical TL element within a bar-coded slide for (Hp [10]) and (Hp [0.07]) was determined and a graph of dose (mGy) against TL signal (Coulomb) was plotted to determine the elements CF. CVs from the raw data for 40 TL elements for Hp (10) and Hp (0.07) were 14.6% and 15.02%, respectively. Further selection of sensitive TL elements reduced the CVs of Hp (10) and Hp (0.07) to 3.73% and 3.21%, respectively, which was seen to be within ±10% accepted limit. The maximum percentage deviation for the calculated and actual dose for Hp (10) and Hp (0.07) was 16.7% and 14.3%, respectively. The CFs were power of 10 − 6 and the Coefficient of determination (R2) for Hp (10) and Hp (0.07) was 0.9998 and 0.9981, respectively, after adjustments were made on the initial graphs. Although large deviations were observed at low doses from the results of the raw data. Re-selected “golden Chips” had R2 close to unity and CV was within recommended standards.


[FULL TEXT] [PDF]*
Print this article     Email this article
Next article
Previous article
Table of Contents
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Citation Manager
Access Statistics
Reader Comments
Email Alert *
Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed2122    
    Printed62    
    Emailed0    
    PDF Downloaded186    
    Comments [Add]    
    Cited by others 2    

Recommend this journal