Home About us Editorial board Search Ahead of print Current issue Archives Submit article Instructions Subscribe Contacts Login 
Home Print this page Email this page Small font size Default font size Increase font size Users Online: 314
ORIGINAL ARTICLE
Year : 2016  |  Volume : 39  |  Issue : 2  |  Page : 96-100

Calibration of Phoswich detector for the measurement of natural uranium in lungs


Radiological Safety Division, Electronics Instrumentation and Radiological Safety Group, Indira Gandhi Center for Atomic Research, Kalpakkam, Tamil Nadu, India

Correspondence Address:
M Manohari
Radiological Safety Division, Electronics Instrumentation and Radiological Safety Group, Indira Gandhi Center for Atomic Research, Kalpakkam, Tamil Nadu
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0972-0464.190391

Rights and Permissions

The occupational workers of fuel fabrication and reprocessing facilities at Indira Gandhi Center for Atomic Research (IGCAR) have a potential for internal exposure to natural uranium which is hazardous both in chemical and radiological aspects. Hence, in vivo monitoring of the radiation workers has to be carried out to ensure safe working conditions. In IGCAR, the in vivo monitoring of natural uranium is being carried out using Phoswich-based lung monitor. The measurement and quantification of internal exposure due to natural uranium is done using 63 and 93 keV photons emitted by 234 Th, immediate daughter of 238 U. Realistic anthropomorphic Lawrence Livermore National Laboratory (LLNL) phantom is used for the calibration of the system. As natural uranium-loaded lung set is not available currently in the laboratory, lung set loaded with indigenously prepared natural uranium source capsules were used. Efficiency curve for 238 U was established for varying muscle equivalent-chest wall thickness (MEQ-CWT) and the efficiency values were in the range of 5.604E-03 to 8.601E-03 CPS/Bq. Simulation results of LLNL voxel phantom having uniform lung distribution of natural uranium agreed within 9% with the measured one which is comparable with the error associated with measurement. This confirms that the distribution pattern of 12 capsules in each lung in the given geometry closely resembles the uniform distribution. The efficiency and minimum detectable activity (MDA) values for Indian population were found to be ranging from 1.011E-03 to 7.931E-03 CPS/Bq and 12–18 Bq, respectively. The efficiency value for 238 U established from the measurement using Japan Atomic Energy Research Institute (JAERI) phantom having uniform source distribution agreed with that of LLNL phantom measurement having capsule source distribution for the same MEQ-CWT thickness (33.8 mm) within 3%. This further reaffirms that the adopted capsule distribution is close to uniform distribution.


[FULL TEXT] [PDF]*
Print this article     Email this article
Next article
Previous article
Table of Contents
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Citation Manager
Access Statistics
Reader Comments
Email Alert *
Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed1224    
    Printed40    
    Emailed0    
    PDF Downloaded168    
    Comments [Add]    
    Cited by others 1    